MMW-05 - Microwave Brain Fixation NEW 5KW
The Muromachi Microwave Brain Fixation system can deactivate brain enzymes in less than 1 second.
Undesired post-mortem modifications in the biochemical composition of the brain can be dramatically reduced by Microwave Brain Fixation.
The Muromachi Microwave Brain Fixation system can deactivate brain enzymes in less than 1 second.
Undesired post-mortem modifications in the biochemical composition of the brain can be dramatically reduced by Microwave Brain Fixation.
The NEW 5 KW machine mantains the same features as the previous model, plus the following new ones:
- Instrument control through the Touch Panel Display on the front panel
- CE-certified
- Air cooling (no circulating water is required)
- Switching power supply to comply with line voltage fluctuation (380-440 V, 20 amp)
The Microwave Brain Fixation system is commonly used prior to analysis of:
• Phosphorylated proteins
• Acetylcholine, Serotonin, Endorphins
• Prostaglandins, Catecholamines
• C-AMP, C-GMP, GABA, DOPA
Read More
Close
Features and Benefits
Specifications
Accessories
Pictures
| NEW FEATURE |
NEW BENEFIT |
| Touch-screen |
Improved usability |
| Air-cooled |
No water circulation required |
| CE-certified |
No additional declaration required |
| Oscillator |
Panasonic Magnetron 2M265-M11 |
| Microwave Power Output |
5kW max.( 2-5 kW variable) |
| Oscillation Frequency |
2,450±30MHz |
| Cooling Requirements |
Air Cooling |
| Exposure Time |
0.10-2.99sec (0.01sec steps) |
| Pausing Time |
3 min |
| Power Consumption |
8kVA |
| Frequency |
50 or 60 Hz |
| Power Requirements |
190-240V 30A 3Phase or
380-440V 20A 3Phase |
| Outside Dimensions |
750W x 550D x 1,275 mmH |
| Weight |
Approx. 103kg |
| APPLICATOR HEAD |
SUITABLE ANIMAL HOLDER |
| TAW-174P |
WJM-24 for 15-20g Mouse
WJM-28 for 20-40g Mouse |
| TAW-424SP |
WJR-S for 150-250g Rat |
| TAW-424MP |
WJR-M for 250-400g Rat
WJR-L for 400-500g Rat |
-
Rodent Brain after in-vivo Microwave Fixation
-
Animal Holders